Infrared gas sensing, answered

Straight answers to the questions we are asked most about NDIR, laser and photoacoustic gas measurement, from choosing a wavelength to understanding what limits a detector.

What is nondispersive infrared (NDIR) gas sensing?

NDIR gas sensing measures how much infrared light a gas absorbs. A source sends light through a sample, a detector sits at the far end, and an optical filter limits the detector to the narrow band of wavelengths that the target gas absorbs. The more gas in the path, the less light arrives. Nondispersive means no prism or grating spreads the light into a spectrum, so the instrument can be small, robust and inexpensive compared with a full spectrometer.

Why does each gas need its own wavelength?

Every molecule absorbs infrared light at wavelengths set by how it vibrates and rotates, so each gas has a characteristic pattern of absorption bands. Methane is commonly measured near 3.31 µm, carbon dioxide near 4.26 µm, nitrous oxide near 4.50 µm, carbon monoxide near 4.67 µm and sulphur dioxide near 7.34 µm. Choosing a band where the target gas absorbs strongly and little else does is the first design decision, and it shapes the choice of source, filter and detector that follow.

What is the Beer Lambert law and why does it matter?

The Beer Lambert law links absorbance to the amount of gas: absorbance equals the absorption strength of the gas at that wavelength, multiplied by its concentration and by the length of the optical path. Transmission, the fraction of light that gets through, falls by a factor of ten for every unit of absorbance. It matters because it lets an engineer predict a signal before any hardware exists. It also shows the limits of a design, since once absorbance reaches about 8 the beam is so strongly absorbed that very little light is left to measure.

What sets the detection limit of an infrared gas sensor?

The detection limit is the smallest concentration that produces a signal larger than the noise. It depends on three things: how strongly the gas absorbs at the chosen wavelength, how long the optical path is, and how quiet the receiver is, usually expressed as a noise equivalent absorbance. A stronger absorber, a longer path or a quieter receiver each lower the limit. Because the three trade against size, cost and power, the useful question is rarely how low can it go, and more often how low does this application need it to go.

How long should the optical path be?

Long enough to produce a measurable absorbance at the lowest concentration of interest, and short enough that the strongest expected concentration does not absorb nearly all of the light. Trace gas measurements often use long folded paths or multipass cells, while measurements of high concentrations can use paths of a few millimetres. The right length follows from the gas, the concentration range and the noise of the receiver, which is why it is worth calculating before a cell is designed.

Why does water vapour cause false readings?

Water vapour absorbs infrared light across wide regions of the spectrum and is nearly always present in air, often at much higher concentrations than the gas being measured. If a target band overlaps part of a water band, a change in humidity can look like a change in the target gas. This is called cross sensitivity. It is handled by choosing a band that avoids water absorption where possible, by using a reference channel, and by correcting for temperature and humidity.

How do laser, NDIR and photoacoustic methods differ?

NDIR typically uses a broadband source and a filter, which keeps it simple and economical. Laser methods use a narrow line source tuned to a single absorption feature, which can give very high selectivity and sensitivity. Photoacoustic methods detect the sound produced when absorbed light heats a gas in a sealed volume, and can reach low detection limits in compact designs. None is best in every case. The right choice depends on the target gas, the required sensitivity, the environment, the size and power available and the cost at volume.

Why model an instrument before building hardware?

Absorption, noise and cross sensitivity can all be calculated, so most of the risk in an infrared design can be examined on a computer before money is spent on optics and electronics. Modelling shows whether a measurement is feasible at all, which wavelength and path length to choose, and how a given detector and source will perform. The same model then becomes the specification for the hardware, and gives something to check the finished instrument against.

What does Scytronix deliver on an infrared project?

Scytronix provides a design service for laser, NDIR, photoacoustic and other infrared chemical measurement systems, supported by bespoke modelling tools. The work typically runs from a feasibility question, through predictive modelling and specification, to a working instrument that answers it. The scope of each engagement is agreed with the client at the start.

What should I bring to a first conversation?

The most useful starting points are the gas or gases to be measured, the range of concentrations expected, the conditions the instrument will work in, such as temperature, humidity and pressure, and any limits on size, power or cost. A rough idea of how the measurement will be used is enough. Everything else can be worked out together.